AD5934YRSZ

Analog Devices — Analog & Mixed-Signal ICs

Product Overview

The AD5934YRSZ from Analog Devices is a high-performance analog / mixed-signal IC providing precision signal acquisition, conversion, and conditioning. Analog Devices (ADI) is a world leader in high-performance analog, mixed-signal, and DSP ICs for instrumentation, healthcare, automotive, and communications.

Superb Automation partners with leading brands including Qorvo, TI, ADI, XILINX, NVIDIA, Intel, ST, MACOM, Renesas, and Coilcraft to provide complete BOM and supply-chain solutions.

Key Features

  • High-resolution data conversion with excellent SNR and THD
  • Wide dynamic range and high linearity across operating conditions
  • Integrated voltage reference with low temperature drift
  • Flexible digital interface — SPI, I2C, or parallel options
  • Industrial/automotive temperature range with guaranteed specs
  • Low power consumption with power-down modes

Key Specifications

Resolutionbits
Channels—
Sample RatekSPS
Interface—
Supply VoltageV
Operating Temp°C
Package—
INLLSB
SNRdB
THDdB

Applications

Precision Instrumentation Industrial Process Control Medical Imaging & Diagnostics Automotive Sensor Interfaces Communications Infrastructure Test & Measurement

Request Quote — In Stock, 100% New & Original, Full Warranty

AD5934YRSZ: A 12-Bit Impedance Converter With a 27-Bit Frequency Word

The AD5934 is a complete impedance measurement system on one die: a frequency generator that excites an external complex impedance with a programmable peak to peak voltage, a 12-bit 250 kSPS analog to digital converter that samples the response, and a digital signal processing engine that performs a discrete Fourier transform at each excitation frequency and returns a real and an imaginary data word. The excitation frequency tops out at 100 kHz, the tuning word is 27 bits wide, which is finer than 0.1 Hz, and the measurement range is 1 kohm to 10 Mohm with the standard circuit, extending down to 100 ohm with additional front end circuitry. System accuracy is 0.5 percent, the supply range is 2.7 V to 5.5 V, and the package is a 16-lead SSOP over minus 40 to plus 125 degrees C.

The 27-bit tuning word is finer than the crystal

The frequency word is 27 bits, which against a 16.776 MHz master clock gives a step of 0.125 Hz, and the data sheet states that resolution below 0.1 Hz is achievable with the right clock. That precision is not useful for measuring an impedance, where 1 Hz is plenty, but it is useful for a sweep that has to land exactly on a mains harmonic or on a resonance that moves with temperature. The practical limit is the master clock accuracy, not the word length: a 50 ppm crystal is a 5 Hz uncertainty at 100 kHz, which is forty times the tuning step.

0.5 percent system accuracy depends on the calibration resistor

The 0.5 percent figure is a system figure, and the system includes the calibration resistor that sets the gain factor at each frequency. A 1 percent resistor in that position makes the reading 1 percent wrong no matter how good the converter is, and a 0.1 percent resistor keeps the converter as the limit. The other term is the output amplifier's finite output resistance, which is why the 1 kohm to 10 Mohm range is the specified one and the 100 ohm to 1 kohm range needs an external buffer: at low impedance the series resistance of the drive path is a significant fraction of the unknown.

ParameterValueWhat it decides
ADC12 bits, 250 kSPSSample rate per frequency point before averaging
Excitation rangeUp to 100 kHzThe top of the sweep, limited by the internal amplifier
Frequency word27 bits0.125 Hz steps at a 16.776 MHz clock
Impedance range1 kohm to 10 MohmThe specified range with the standard front end
Extended range100 ohm to 1 kohmNeeds additional circuitry at the drive output
System accuracy0.5 percentIncludes the calibration resistor tolerance
Excitation amplitudeProgrammable peak to peakSets the signal to noise ratio against the unknown
OutputReal and imaginary words per pointMagnitude and phase follow by arithmetic
Supply2.7 V to 5.5 VPortable instruments and 5 V racks from one part
InterfaceI2CTwo wire bus, address selected by pin
ClockExternal master clock16.776 MHz is the value the timing assumes
Package16-lead SSOPMinus 40 to plus 125 degrees C
Related partAD5933, 1 MSPS with a temperature sensorFaster conversion, same measurement method

What the DFT engine actually returns

At each frequency the part multiplies the sampled response by a reference sine and cosine and accumulates the products, which is a single bin discrete Fourier transform. The result is a real and an imaginary word proportional to the admittance of the unknown, so the magnitude is the square root of the sum of the squares and the phase is the arctangent of their ratio. Everything else, including the calibration, happens in the host. That division of labour is why the part is cheap and why it needs a calibration procedure: the converter produces numbers, not ohms.

Choosing the excitation amplitude

The excitation voltage is programmable, and the choice trades signal to noise against linearity in the unknown. A high amplitude gives a better signal to noise ratio at the converter, but a biological or electrochemical sample is nonlinear and a large excitation changes the very impedance being measured. The usual approach is to sweep at the amplitude the application will actually see, accept the noise, and average more points. Averaging reduces the random term and does nothing for the nonlinear one, so the amplitude is a measurement definition rather than a setting to be optimized for the instrument.

Sourcing an impedance measurement front end

This class of part is bought against a measurement method, so the substitute question is the excitation range and the frequency word, not the pinout. Parts that combine a converter with a signal generator and can be compared on those terms are listed in ADC DAC cross reference, the medical instrument side of the same family of parts is covered in medical device components, and the supply route for a small instrument bill of materials is in one stop BOM sourcing.