Contact Us
  • Home
  • COMPONENTS
  • Particulate Matter Sensor Modules for Air Quality Monitors

Particulate Matter Sensor Modules for Air Quality Monitors

A particulate matter sensor counts and weighs what floats in the air, and the channel list it reports is as important as the accuracy. The modules below use laser scattering with a fan to pull a known volume past the beam, so the reading is a mass concentration in micrograms per cubic metre. One reports six size bins from 0.3 to 10 micrometre, the other reports PM1, PM2.5, PM4 and PM10 plus a number concentration for each. Both are quoted with the sample volume, the response time and the laser class, because those three decide how the reading behaves in a duct.
quote now

Product Specifications

Product Overview

A laser scattering particle sensor pulls a small volume of air through a beam and counts the flashes. The count becomes a mass concentration through an assumed particle density and an assumed size distribution, and that is why two sensors on the same air can differ by a factor: they are solving different sums on the same light. Reading the output format before the accuracy is the difference between a monitor that agrees with a reference and one that does not.

The first figure to check is the sample volume and the response time. A 0.1 litre sample updated every second is a different measurement from a 1 second integration over a fan driven path, and in a duct with a changing concentration the two will not track each other. The second is the size bins, because PM2.5 is defined by a cut at 2.5 micrometre and a sensor that reports a cumulative count above a threshold is not reporting the same quantity as one that bins the sizes.

The third figure is the laser class and the airflow path. A class 1 laser in a sealed optical bench is safe in a wall sensor, while a sensor that needs a fan and an open path has to be mounted so that dust does not settle on the window. The fourth is calibration: these modules are calibrated against a reference instrument with a particular aerosol, and a monitor used for wood smoke or for salt spray will read differently unless it is corrected in the field.

The fifth point is what the host does with the data. A room monitor needs a PM2.5 number once a minute, an occupational monitor needs both mass and number, and a filter monitor needs the trend rather than the absolute value. The module output format decides which of those is possible without extra processing, and it is the first line to compare when two parts look similar on the front page.

Key Specifications

ParameterValueNote
Measurement principlelaser scattering with a fan driven samplethe count becomes mass through an assumed density and distribution
Mass concentration range0 to 500 micrograms per cubic metre effective on the 0.1 litre module, 0 to 1,000 on the fan driven modulethe maximum range is above the effective range
Size channels0.3 to 1.0, 1.0 to 2.5 and 2.5 to 10 micrometre on one module, PM1, PM2.5, PM4 and PM10 on the othera cumulative count is not the same quantity as a binned mass
Resolution1 microgram per cubic metre on the 0.1 litre modulethe smallest step the output can show
Accuracyplus or minus 10 percent between 100 and 500 micrograms, plus or minus 10 micrograms below 100stated against a reference aerosol, not against any dust
Sample volume0.1 litre standard volume on the 0.1 litre modulethe sample volume sets the response time
Response timeunder 1 s per sample and 10 s total on the 0.1 litre module, 1 s sampling interval on the fan driven modulea duct measurement needs the faster part
Supply voltage5 V, 4.5 to 5.5 V on the 0.1 litre module, 4.5 to 5.5 V on the fan driven moduleboth parts want a 5 V rail with headroom
Currentunder 100 mA active and 200 microampere standby on the 0.1 litre module, 55 mA typical and 80 mA peak on the fan driven modulethe fan start is the peak, not the average
InterfaceUART on the 0.1 litre module, UART at 115200 baud or I2C at address 0x69 on the fan driven moduleI2C needs a cable under 10 cm
Laserclass 1 in both parts, 660 nm on the fan driven modulethe optical bench is sealed
Operating rangeminus 10 to plus 60 C and 0 to 99 percent RH on the 0.1 litre module, minus 10 to plus 60 C and 0 to 95 percent RH on the fan driven modulehumidity above 80 percent affects both
Lifetimeover 10 years of continuous operation on the fan driven module, 3 years minimum on the 0.1 litre modulethe fan and the window are what age
Size50 by 38 by 21 mm on the 0.1 litre module, 41 by 41 by 12 mm on the fan driven modulethe smaller part is the one that fits a wall plate

Included Models and Order Codes

Two modules cover most monitor builds. The 0.1 litre module is the widely used fan driven sensor with six size bins, a 1 microgram resolution and a UART output, and it is the part that most low cost monitors are built around. The fan driven module from the second maker adds number concentration for PM0.5 to PM10, a 660 nm laser in a sealed bench, a 41 by 41 mm footprint and a lifetime above ten years, and it offers I2C as well as UART. Both are calibrated against a reference instrument with a specific aerosol, so the field correction is part of the build, not an afterthought.

ModuleChannelsRangeOutputSupplyNote
PMS50030.3 to 1.0, 1.0 to 2.5, 2.5 to 10 micrometre, six cumulative counts0 to 500 micrograms per cubic metre effective, above 1,000 maximumUART, 1 microgram resolution5 V, under 100 mA0.1 litre sample, 50 by 38 by 21 mm
SPS30PM1, PM2.5, PM4, PM10 mass plus number concentration for each0 to 1,000 micrograms per cubic metre, 0 to 3,000 per cubic centimetreUART at 115200 baud or I2C at 0x694.5 to 5.5 V, 55 mA typical41 by 41 by 12 mm, 660 nm laser, over 10 years

Reading a Particle Sensor Output

The two modules in the table do not report the same quantity, and that is the usual reason two monitors disagree on the same air. The table is the short version of what each output line means.

Output lineWhat it isWhere it misleadsWhat to do
Cumulative count above 0.3 micrometreparticles larger than a threshold, countedit is a count, not a massconvert with a density and a distribution, or read the mass channel
PM2.5 massmass of particles below 2.5 micrometredepends on the assumed densitycorrect against a reference for the local aerosol
PM10 massmass of particles below 10 micrometrecoarse dust settles in the inletmount the inlet downward and clean it
Number concentrationparticles per cubic centimetre by sizenot comparable to a mass figureuse it for trend and for filter loading
Atmospheric and standard conversiontwo ways of scaling a count to a massthe two differ above 100 microgramsstate which one the monitor reports
Humidity above 80 percentwater droplets scatter like particlesa humid day reads as pollutionheat the inlet or correct for humidity

Key Features

Six size bins on one part – the 0.1 litre module reports particles above 0.3, 0.5, 1.0, 2.5, 5 and 10 micrometre, which covers the channels a monitor normally shows.

Number and mass together – the fan driven module reports both, so a filter loading trend and a mass reading come from one sensor.

Sealed optical bench – a class 1 laser in a sealed path is what lets the smaller module sit behind a wall plate without a light trap.

Sample volume stated – 0.1 litre on one part is written down, because the sample volume is what sets the response time.

Lifetime figure – over ten years on the fan driven module, which is the number a maintenance contract is written against.

Two interfaces – UART at 115200 baud or I2C at 0x69, so the module can sit on an existing bus or on a serial port.

Low standby current – 200 microampere standby on the 0.1 litre module keeps a battery monitor alive between readings.

Peak current stated – the fan start is the peak, and a host that sizes its regulator on the average will brown out at start up.

Applications

Three monitors where the channel list and the airflow decide the module rather than the accuracy figure.

Wall mounted room monitor – a 41 by 41 mm module with PM2.5 and number concentration sits behind a louvre, and the sealed bench means no light trap is needed.

Duct and HVAC monitor – the 0.1 litre module with a 1 s update follows a duct that changes with the damper, where a slower part would average the change away.

Occupational and filter monitoring – number concentration is the line that shows a filter loading, while the mass channel is the line that goes into the report.

A particle monitor is a fan, a laser driver and a serial port, so the SUPERB PCBA and the SUPERB PCB keep the fan current loop away from the serial lines on the same board.

Sourcing and Cross-Reference

The first question on a particle monitor is which quantity the report needs. A mass figure and a number figure answer different questions, and a module that reports only cumulative counts needs a conversion that depends on the aerosol. The second question is the inlet, because a downward facing inlet with a coarse screen keeps dust from settling in the optical path.

The second question is humidity. Above 80 percent relative humidity, water droplets scatter light like particles, and a monitor in a coastal or a kitchen environment will read high on humid days. Heating the inlet by a few degrees, or correcting with a humidity reading, is the usual fix, and both need space in the enclosure.

The third question is calibration against the local aerosol. These modules are calibrated with a reference instrument on a particular dust, and wood smoke, diesel and salt spray each read differently. A field correction factor is normal practice, and the reference instrument used to derive it should be named in the report. See lead time and availability for a specific module and BOM shortage recovery where a build is held up by one part.

The SUPERB Assembly team fits the inlet louvre before the module is seated, because a louvre added later leaves a gap that lets dust settle on the window.

Q: Can you supply the particle module with the connector and cable already matched to our host board, so no adapter is needed?

A: Yes. Send the channels, the enclosure, the supply and the interface, and the SUPERB Tech team returns the modules and the cable that fit.

Why Monitor Builders Source Particle Modules Through SUPERB

  1. Output format stated: mass, number and cumulative count are named per part, so a report is not written against the wrong quantity.
  2. Peak current listed: the fan start current is written down, because a regulator sized on the average browns out at start up.
  3. Lifetime figure: the ten year figure on the fan driven part is what a maintenance interval is planned from.
  4. Inlet guidance: the mounting direction and the humidity limit are given with the part, not left to the field.

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: CO2 Sensor Modules for Medical and Building Use, Gas Sensors for Air Quality and Leak Detection, Gas Mass Flow Sensors for Ventilators and Analysers, Infrared Thermopile Temperature Sensors, Water Quality Sensor Circuits for pH and Conductivity, Component Lead Time Solutions.