ECG and Biopotential Analog Front Ends for Patient Monitors
Product Specifications
Product Overview
A biopotential front end is judged by three numbers and one feature set. The noise floor in the band of interest decides whether a small signal is visible, the input bias current decides how much the electrode impedance disturbs the trace, and the common mode rejection decides whether the reading survives the mains field the patient sits in. The feature set, meaning right leg drive, lead off detection and respiration measurement, decides how many external parts the front end needs before it can be connected to a cable.
This page groups the front ends we supply from the single lead parts up to the eight channel monitor devices. The two channel family runs 24 bit conversion with 8 uVpp input referred noise in a 150 Hz bandwidth at gain 6, 200 pA input bias current, 105 dB common mode rejection, 335 uW per channel and data rates from 125 SPS to 8 kSPS. The eight channel family drops the noise to 4 uVpp at gain 6, raises rejection to 115 dB, keeps the same 200 pA bias current and covers 250 SPS to 32 kSPS at 0.75 mW per channel.
Both families integrate the programmable gain amplifier, the reference and the oscillator, and both provide the lead off detection current settings of 6, 12, 18 and 24 nA, a right leg drive amplifier, Wilson centre terminal and pace detection. The R versions add respiration impedance measurement on the same channel set. For a single lead wearable the small front ends give a complete acquisition chain with integrated filters and amplifier stages in one part, and the front end is normally supplied together with the medical grade passive parts and the isolated supply for the same board.
Key Specifications
| Parameter | Two Channel Family | Eight Channel Family |
|---|---|---|
| Resolution | 24 bit, simultaneous sampling | 24 bit, simultaneous sampling |
| Input referred noise | 8 uVpp in 150 Hz bandwidth at gain 6 | 4 uVpp in 150 Hz bandwidth at gain 6 |
| Input bias current | 200 pA | 200 pA |
| Common mode rejection | 105 dB | 115 dB |
| Data rate | 125 SPS to 8 kSPS | 250 SPS to 32 kSPS |
| Power per channel | 335 uW | 0.75 mW |
| Gain settings | 1, 2, 3, 4, 6, 8, 12 | 1, 2, 3, 4, 6, 8, 12 |
| Supply | Analog 2.7 V to 5.25 V, digital 1.7 V to 3.6 V | Analog 2.7 V to 5.25 V, digital 1.65 V to 3.6 V |
| Integrated features | Right leg drive, lead off detection, pace detection, oscillator and reference | Right leg drive, lead off detection, Wilson centre terminal, pace detection, respiration impedance measurement |
| Interface | SPI compatible serial | SPI compatible serial, devices can be daisy chained |
Included Models and Order Codes
The table lists the front end part codes most often requested for monitors, Holter recorders and wearable patches. Send the complete order code including the package and we confirm availability against that string.
| Part Number | Channels | Noise and Power | Features | Package | Notes |
|---|---|---|---|---|---|
| ADS1291 | 1 channel, 24 bit | Low power single channel front end | PGA, reference, oscillator | Surface mount | Single lead acquisition with the internal reference |
| ADS1292 | 2 channels, 24 bit | 8 uVpp at gain 6, 335 uW per channel | Right leg drive, lead off detection, pace detection | TQFP or QFN | 105 dB common mode rejection, 125 SPS to 8 kSPS |
| ADS1292R | 2 channels, 24 bit with respiration | 8 uVpp at gain 6, 335 uW per channel | Adds respiration impedance measurement to the 1292 feature set | TQFP or QFN | Used on monitors that record breathing from the ECG electrodes |
| ADS1294 | 4 channels, 24 bit | 4 uVpp at gain 6, 0.75 mW per channel | Right leg drive, lead off detection, pace detection | TQFP or QFN | 250 SPS to 32 kSPS |
| ADS1296 | 6 channels, 24 bit | 4 uVpp at gain 6, 0.75 mW per channel | Same feature set as the four channel part | TQFP or QFN | For six lead configurations |
| ADS1298 | 8 channels, 24 bit | 4 uVpp at gain 6, 0.75 mW per channel | Right leg drive, Wilson centre terminal, lead off detection, pace detection, respiration option | TQFP or QFN | Referenced to AAMI EC11, EC13, IEC 60601-1, IEC 60601-2-27 and IEC 60601-2-51 |
| ADS1294R | 4 channels, 24 bit with respiration | 4 uVpp at gain 6 | Adds respiration impedance measurement | TQFP or QFN | Same footprint as the 1294 |
| ADS1296R | 6 channels, 24 bit with respiration | 4 uVpp at gain 6 | Adds respiration impedance measurement | TQFP or QFN | Same footprint as the 1296 |
| ADS1298R | 8 channels, 24 bit with respiration | 4 uVpp at gain 6 | Full monitor feature set with respiration | TQFP or QFN | Daisy chain support for more than eight channels |
| AD8232 | 1 channel, single lead heart rate front end | Integrated filters and amplifier stages | Right leg drive style reference and fast restore | Small surface mount | Used on wearable chest band and fitness modules |
| ADAS1000-4 | 4 channel ECG front end | Low power monitor front end | Pace detection, lead off detection, respiration | Surface mount | Multichannel monitor front end with SPI interface |
Key Features
Noise floor stated per family – 8 uVpp for the two channel parts and 4 uVpp for the eight channel parts, both in a 150 Hz bandwidth at gain 6, which is the number that decides whether the trace is usable
Input bias current quoted – 200 pA on both families, which sets the electrode impedance the front end can tolerate before the trace drifts
Lead off detection built in – current settings of 6, 12, 18 and 24 nA with comparator threshold accuracy of plus or minus 30 mV, so no external detection circuit is added
Right leg drive amplifier on chip – the common mode loop is closed on the front end rather than on a separate amplifier, which simplifies the patient cable
Respiration on the same channels – the R parts measure respiration impedance without an extra electrode pair or a separate carrier circuit
Standards referenced parts – the eight channel front end is documented against AAMI EC11 and EC13 and the IEC 60601 series so the equipment file can reference the part directly
Applications
The parts are supplied for the boards below. The acquisition channel is normally quoted with the SUPERB PCB and SUPERB PCBA work we support for the same board, so the front end, the medical grade passives and the assembly are handled as one order.
Patient monitors – multi channel ECG acquisition with lead off detection, pace detection and respiration in one front end per patient input block
Holter and telemetry recorders – low power two channel front ends for battery operated recording where 335 uW per channel sets the run time
Wearable and remote monitoring – single lead and two channel parts on patch and chest band products, with integrated filters to keep the electrode front end small
Sourcing and Cross-Reference
Medical front ends are the hardest parts to replace. A substitution changes noise, bias current and the internal filter response at the same time, and every one of those has to be re-verified against the equipment file. That is why availability has to be confirmed against the exact order code rather than the family.
We confirm front end part numbers against the complete order code including package, and where an order code is short we quote the nearest option in the same footprint with the noise, data rate and power difference stated in writing, so the verification plan only covers the differences that actually changed. Quantities on long lead time can be scheduled against a production date, and the medical grade passives and isolated supply for the same front end can be quoted on one line.
For volume medical builds, SUPERB Assembly support covers the SMT assembly of the acquisition section.
Q: Can you supply the same order code with a certified date code for a medical build?
A: Yes. Send the complete order code and the quantity and the SUPERB Tech sales team will confirm availability, the batch information and the nearest option in the same footprint.
Why Buy From Us
- Order code and date code confirmed: front ends are quoted against the full order code with batch and date code information for the equipment file
- Differences stated before replacement: noise, data rate and supply differences are listed in writing when a nearest option is quoted
- Whole acquisition section sourced: front end, medical grade passives and isolated supply can be quoted together so the channel is built from one delivery
- Lead time scheduled, not improvised: long lead time quantities are reserved against a production date instead of being chased on the spot market